Type
Competitive ELISA, Immobilized antigen
Applications
Serum
Sample Requirements
50 μl/well
Shipping
At ambient temperature. Upon receipt, store the product at the temperature recommended below.
Storage/Expiration
Store the complete kit at 2–8°C. Under these conditions, the kit is stable until the expiration date (see label on the box).
Calibration Curve
Calibration Range
0.25–50 ng/ml
Limit of Detection
0.1 ng/ml
Intra-assay (Within-Run)
CV = 7.0%
Inter-assay (Run-to-Run)
CV = 9.2%
Spiking Recovery
94,40%
Dilutation Linearity
93,80%