Type
 
											Sandwich ELISA
												Applications
 
											Plasma-EDTA, Urine
												Sample Requirements
 
											50 µl/well
												Shipping
 
											Shipped on blue ice packs. Upon receipt, store the product at the temperature recommended below.
												Storage/Expiration
 
											Store the complete kit at 2–8°C. Under these conditions, all components are stable until the expiration date (see label on the box).
												Calibration Range
 
											50 - 2000 pg/ml
												Limit of Detection
 
											25 pg/ml
												Intra-assay (Within-Run)
 
											n = 8
CV = 4.8%
												Inter-assay (Run-to-Run)
 
											n = 8
CV = 9.2%
												Spiking Recovery
 
											88.0%
												Dilution Linearity
 
											107.3%
								
							 
							
								
												Features
 
											It is intended for research use only
The total assay time is less than 20 hours
The kit measures FGF-23
Assay format is 96 wells
Quality Controls are human serum based
Standard is a recombinant protein
Components of the kit are provided ready to use, concentrated or dried
												Research topic
 
											Bone and cartilage metabolism, Renal disease
												Summary
 
											Fibroblast growth factor-23 (FGF-23) is a phosphaturic hormone involved in mineral bone metabolism that helps control phosphate homeostasis and reduces 1,25- dihydroxyvitamin D synthesis. Recent data have highlighted the relevant direct FGF-23 effects on the myocardium, and high plasma levels of FGF-23 have been associated with adverse cardiovascular outcomes in humans, such as heart failure and arrhythmias. Therefore, FGF-23 has emerged as a novel biomarker of cardiovascular risk. experimental data suggest FGF-23 as a direct mediator of cardiac hypertrophy development, cardiac fibrosis and cardiac dysfunction via specific myocardial FGF receptor (FGFR) activation.
								
							 
							
	
				Find documents for the lot
				
				
						
						Example Instructions for Use (RUO)
						
					
							Example Instructions for Use (RUO)